| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 2000.00 | 2.061 | 1.979 |
| 1000.00 | 1.643 | 1.561 |
| 500.00 | 1.214 | 1.132 |
| 250.00 | 0.713 | 0.631 |
| 125.00 | 0.629 | 0.547 |
| 62.50 | 0.335 | 0.253 |
| 31.25 | 0.220 | 0.138 |
| 0.00 | 0.082 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 83-113% | 98% |
| EDTA plasma(n=5) | 83-121% | 102% |
| Heparin plasma(n=5) | 89-120% | 105% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 91-99% | 89-106% | 90-114% | 88-114% |
| EDTA plasma(n=5) | 81-102% | 89-105% | 82-106% | 86-103% |
| Heparin plasma(n=5) | 83-102% | 87-113% | 94-102% | 91-109% |