| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 5000.00 | 2.857 | 2.764 |
| 2500.00 | 2.244 | 2.151 |
| 1250.00 | 1.289 | 1.196 |
| 625.00 | 0.616 | 0.523 |
| 312.50 | 0.334 | 0.241 |
| 156.25 | 0.205 | 0.112 |
| 78.13 | 0.108 | 0.015 |
| 0.00 | 0.093 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 84-94% | 89% |
| EDTA plasma(n=5) | 90-104% | 97% |
| Heparin plasma(n=5) | 94-105% | 100% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 86-97% | 78-89% | 78-92% | 92-105% |
| EDTA plasma(n=5) | 91-97% | 78-91% | 82-88% | 82-88% |
| Heparin plasma(n=5) | 88-94% | 89-99% | 92-106% | 93-99% |