| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 2.241 | 2.155 |
| 500.00 | 1.744 | 1.658 |
| 250.00 | 1.044 | 0.958 |
| 125.00 | 0.859 | 0.773 |
| 62.50 | 0.534 | 0.448 |
| 31.25 | 0.342 | 0.256 |
| 15.63 | 0.229 | 0.143 |
| 0.00 | 0.086 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 94-100% | 97% |
| EDTA plasma(n=5) | 89-120% | 105% |
| Heparin plasma(n=5) | 81-86% | 84% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 89-113% | 95-109% | 83-99% | 92-103% |
| EDTA plasma(n=5) | 89-106% | 81-116% | 86-101% | 84-95% |
| Heparin plasma(n=5) | 83-98% | 86-113% | 91-107% | 80-94% |