| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 1.970 | 1.883 |
| 500.00 | 1.673 | 1.586 |
| 250.00 | 1.130 | 1.043 |
| 125.00 | 0.800 | 0.713 |
| 62.50 | 0.531 | 0.444 |
| 31.25 | 0.374 | 0.287 |
| 15.63 | 0.227 | 0.140 |
| 0.00 | 0.087 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 95-107% | 101% |
| EDTA plasma(n=5) | 91-113% | 102% |
| Heparin plasma(n=5) | 90-105% | 98% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 84-118% | 89-116% | 85-109% | 90-111% |
| EDTA plasma(n=5) | 84-111% | 85-118% | 86-119% | 90-105% |
| Heparin plasma(n=5) | 82-119% | 86-102% | 92-103% | 90-107% |