| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 2000.00 | 2.201 | 2.112 |
| 1000.00 | 1.593 | 1.504 |
| 500.00 | 1.135 | 1.046 |
| 250.00 | 0.933 | 0.844 |
| 125.00 | 0.591 | 0.502 |
| 62.50 | 0.363 | 0.274 |
| 31.25 | 0.196 | 0.107 |
| 0.00 | 0.089 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 94-100% | 97% |
| EDTA plasma(n=5) | 77-91% | 84% |
| Heparin plasma(n=5) | 89-95% | 92% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 93-99% | 78-84% | 81-89% | 79-87% |
| EDTA plasma(n=5) | 93-105% | 90-97% | 82-96% | 82-96% |
| Heparin plasma(n=5) | 77-88% | 88-95% | 84-98% | 86-100% |