| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (μg/mL) | OD | Corrected OD |
|---|---|---|
| 500.00 | 0.179 | 0.000 |
| 250.00 | 0.397 | 0.000 |
| 125.00 | 0.722 | 0.000 |
| 62.50 | 0.993 | 0.000 |
| 31.25 | 1.236 | 0.000 |
| 15.63 | 1.589 | 0.000 |
| 7.82 | 1.869 | 0.000 |
| 0.00 | 2.232 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 84-94% | 89% |
| EDTA plasma(n=5) | 82-88% | 85% |
| Heparin plasma(n=5) | 80-89% | 85% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 84-94% | 78-85% | 89-102% | 82-95% |
| EDTA plasma(n=5) | 78-88% | 92-104% | 87-93% | 87-93% |
| Heparin plasma(n=5) | 77-84% | 80-92% | 88-98% | 83-89% |