| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 5000.00 | 2.311 | 2.242 |
| 2500.00 | 1.784 | 1.715 |
| 1250.00 | 1.216 | 1.147 |
| 625.00 | 0.834 | 0.765 |
| 312.50 | 0.492 | 0.423 |
| 156.25 | 0.317 | 0.248 |
| 78.13 | 0.236 | 0.167 |
| 0.00 | 0.069 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 77-89% | 83% |
| EDTA plasma(n=5) | 94-104% | 99% |
| Heparin plasma(n=5) | 78-84% | 81% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 84-98% | 81-88% | 77-86% | 94-107% |
| EDTA plasma(n=5) | 89-101% | 78-91% | 94-106% | 94-106% |
| Heparin plasma(n=5) | 84-96% | 86-99% | 77-83% | 93-100% |