| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 300.00 | 0.191 | 0.000 |
| 150.00 | 0.386 | 0.000 |
| 75.00 | 0.655 | 0.000 |
| 37.50 | 0.878 | 0.000 |
| 18.75 | 1.139 | 0.000 |
| 9.38 | 1.578 | 0.000 |
| 4.69 | 1.796 | 0.000 |
| 0.00 | 2.232 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 86-94% | 90% |
| EDTA plasma(n=5) | 87-95% | 91% |
| Heparin plasma(n=5) | 86-97% | 92% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 77-91% | 93-99% | 90-98% | 95-102% |
| EDTA plasma(n=5) | 85-92% | 84-97% | 94-106% | 94-106% |
| Heparin plasma(n=5) | 83-90% | 83-94% | 86-93% | 95-106% |