| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 5000.00 | 2.302 | 2.219 |
| 2500.00 | 1.777 | 1.694 |
| 1250.00 | 1.226 | 1.143 |
| 625.00 | 0.839 | 0.756 |
| 312.50 | 0.519 | 0.436 |
| 156.25 | 0.304 | 0.221 |
| 78.13 | 0.196 | 0.113 |
| 0.00 | 0.083 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 78-87% | 83% |
| EDTA plasma(n=5) | 87-99% | 93% |
| Heparin plasma(n=5) | 82-96% | 89% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 83-93% | 96-103% | 92-100% | 94-100% |
| EDTA plasma(n=5) | 89-97% | 91-103% | 90-96% | 90-96% |
| Heparin plasma(n=5) | 87-96% | 95-107% | 78-88% | 93-105% |