| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 5000.00 | 2.168 | 2.080 |
| 2500.00 | 1.636 | 1.548 |
| 1250.00 | 1.224 | 1.136 |
| 625.00 | 0.799 | 0.711 |
| 312.50 | 0.536 | 0.448 |
| 156.25 | 0.358 | 0.270 |
| 78.13 | 0.256 | 0.168 |
| 0.00 | 0.088 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 96-105% | 101% |
| EDTA plasma(n=5) | 95-103% | 99% |
| Heparin plasma(n=5) | 86-94% | 90% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 96-107% | 87-96% | 89-98% | 90-104% |
| EDTA plasma(n=5) | 89-99% | 95-101% | 89-102% | 89-102% |
| Heparin plasma(n=5) | 90-96% | 79-88% | 95-105% | 88-95% |