| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 2000.00 | 1.936 | 1.853 |
| 1000.00 | 1.645 | 1.562 |
| 500.00 | 1.251 | 1.168 |
| 250.00 | 0.916 | 0.833 |
| 125.00 | 0.578 | 0.495 |
| 62.50 | 0.334 | 0.251 |
| 31.25 | 0.249 | 0.166 |
| 0.00 | 0.083 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 84-102% | 93% |
| EDTA plasma(n=5) | 82-120% | 101% |
| Heparin plasma(n=5) | 89-104% | 97% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 91-102% | 91-102% | 83-113% | 89-105% |
| EDTA plasma(n=5) | 90-99% | 86-104% | 86-98% | 82-111% |
| Heparin plasma(n=5) | 90-98% | 88-105% | 87-119% | 84-114% |