| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 500.00 | 0.221 | |
| 250.00 | 0.411 | |
| 125.00 | 0.715 | |
| 62.50 | 0.933 | |
| 31.25 | 1.225 | |
| 15.63 | 1.587 | |
| 7.82 | 1.761 | |
| 0.00 | 2.211 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 95-107% | 101% |
| EDTA plasma(n=5) | 80-92% | 86% |
| Heparin plasma(n=5) | 95-107% | 101% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 89-95% | 93-102% | 85-92% | 90-100% |
| EDTA plasma(n=5) | 88-98% | 79-88% | 77-90% | 77-90% |
| Heparin plasma(n=5) | 93-100% | 93-104% | 96-107% | 81-90% |