| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 2.010 | 1.936 |
| 500.00 | 1.570 | 1.496 |
| 250.00 | 1.142 | 1.068 |
| 125.00 | 0.740 | 0.666 |
| 62.50 | 0.428 | 0.354 |
| 31.25 | 0.380 | 0.306 |
| 15.63 | 0.220 | 0.146 |
| 0.00 | 0.074 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 93-115% | 104% |
| EDTA plasma(n=5) | 83-128% | 106% |
| Heparin plasma(n=5) | 82-103% | 93% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 85-120% | 87-115% | 91-106% | 90-99% |
| EDTA plasma(n=5) | 94-103% | 83-115% | 84-119% | 89-96% |
| Heparin plasma(n=5) | 83-101% | 91-116% | 93-118% | 95-109% |