Products

Fish TC(Total cholesterol) ELISA Kit
One-click to copy product information
ELK11735
Size: Price:
48 T $412.00
96 T $588.00
96 T*5 $2499.00

Overview

Product name: Fish TC(Total cholesterol) ELISA Kit
Reactivity: Fish
Alternative Names: TC
Assay Type: Competitive Inhibition
Sensitivity: 2.64 μg/mL
Standard: 500 μg/mL
Detection Range: 7.82-500 μg/mL
Sample Type: serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids
Assay Length: 2h
Research Area: Metabolic pathway;Hepatology;Gastroenterology;Nutrition metabolism;Hormone metabolism;
Test principle: This assay employs the competitive inhibition enzyme immunoassay technique. The microtiter plate provided in this kit has been pre-coated with Fish TC. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Fish TC. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Fish TC in the samples is then determined by comparing the OD of the samples to the standard curve.

Standard curve

Concentration (μg/mL) OD Corrected OD
500.00 0.191
250.00 0.336
125.00 0.524
62.50 0.831
31.25 1.289
15.63 1.501
7.82 1.786
0.00 2.236

Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

Recovery

Matrices listed below were spiked with certain level of recombinant TC and the recovery rates were calculated by comparing the measured value to the expected amount of TC in samples.
Matrix Recovery range Average
serum(n=5) 83-95% 89%
EDTA plasma(n=5) 80-96% 88%
Heparin plasma(n=5) 80-95% 87%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of TC and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 86-93% 88-92% 86-97% 90-101%
EDTA plasma(n=5) 91-99% 85-93% 82-95% 93-99%
Heparin plasma(n=5) 82-94% 85-96% 78-90% 91-99%