| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 500.00 | 2.368 | 2.282 |
| 250.00 | 1.608 | 1.522 |
| 125.00 | 1.223 | 1.137 |
| 62.50 | 0.847 | 0.761 |
| 31.25 | 0.593 | 0.507 |
| 15.63 | 0.363 | 0.277 |
| 7.81 | 0.194 | 0.108 |
| 0.00 | 0.086 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 89-103% | 96% |
| EDTA plasma(n=5) | 83-99% | 95% |
| Heparin plasma(n=5) | 81-95% | 88% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 96-101% | 95-104% | 79-92% | 82-95% |
| EDTA plasma(n=5) | 89-103% | 85-93% | 89-97% | 95-102% |
| Heparin plasma(n=5) | 90-99% | 95-102% | 82-90% | 85-92% |