| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 2000.00 | 2.332 | 2.246 |
| 1000.00 | 1.633 | 1.547 |
| 500.00 | 1.296 | 1.210 |
| 250.00 | 0.876 | 0.790 |
| 125.00 | 0.543 | 0.457 |
| 62.50 | 0.315 | 0.229 |
| 31.25 | 0.189 | 0.103 |
| 0.00 | 0.086 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 80-96% | 88% |
| EDTA plasma(n=5) | 82-98% | 90% |
| Heparin plasma(n=5) | 83-97% | 90% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 95-102% | 92-103% | 83-98% | 88-97% |
| EDTA plasma(n=5) | 92-96% | 86-98% | 95-101% | 87-94% |
| Heparin plasma(n=5) | 87-96% | 89-98% | 83-103% | 96-105% |