| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 40.00 | 2.127 | 2.034 |
| 20.00 | 1.554 | 1.461 |
| 10.00 | 1.048 | 0.955 |
| 5.00 | 0.987 | 0.894 |
| 2.50 | 0.580 | 0.487 |
| 1.25 | 0.385 | 0.292 |
| 0.63 | 0.198 | 0.105 |
| 0.00 | 0.093 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 94-107% | 101% |
| EDTA plasma(n=5) | 90-117% | 104% |
| Heparin plasma(n=5) | 81-91% | 86% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 87-115% | 90-104% | 83-118% | 86-106% |
| EDTA plasma(n=5) | 91-105% | 86-118% | 92-98% | 83-105% |
| Heparin plasma(n=5) | 84-113% | 82-88% | 91-112% | 91-102% |