| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 2.328 | 2.239 |
| 500.00 | 1.616 | 1.527 |
| 250.00 | 1.269 | 1.180 |
| 125.00 | 0.869 | 0.780 |
| 62.50 | 0.539 | 0.450 |
| 31.25 | 0.352 | 0.263 |
| 15.63 | 0.189 | 0.100 |
| 0.00 | 0.089 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 87-99% | 93% |
| EDTA plasma(n=5) | 80-95% | 87% |
| Heparin plasma(n=5) | 78-90% | 84% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 82-96% | 86-97% | 83-96% | 91-103% |
| EDTA plasma(n=5) | 95-104% | 87-96% | 89-101% | 82-90% |
| Heparin plasma(n=5) | 85-95% | 89-99% | 85-95% | 87-96% |