| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 500.00 | 2.354 | 2.269 |
| 250.00 | 1.594 | 1.509 |
| 125.00 | 1.252 | 1.167 |
| 62.50 | 0.866 | 0.781 |
| 31.25 | 0.567 | 0.482 |
| 15.63 | 0.312 | 0.227 |
| 7.81 | 0.215 | 0.130 |
| 0.00 | 0.085 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 87-93% | 90% |
| EDTA plasma(n=5) | 85-97% | 91% |
| Heparin plasma(n=5) | 95-105% | 100% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 93-102% | 90-99% | 95-106% | 87-101% |
| EDTA plasma(n=5) | 85-98% | 97-104% | 92-101% | 82-96% |
| Heparin plasma(n=5) | 79-96% | 89-100% | 85-96% | 86-99% |