| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 2.345 | 2.280 |
| 500.00 | 1.621 | 1.556 |
| 250.00 | 1.224 | 1.159 |
| 125.00 | 0.828 | 0.763 |
| 62.50 | 0.545 | 0.480 |
| 31.25 | 0.236 | 0.171 |
| 15.63 | 0.232 | 0.167 |
| 0.00 | 0.065 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 89-103% | 96% |
| EDTA plasma(n=5) | 90-105% | 97% |
| Heparin plasma(n=5) | 88-102% | 95% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 85-96% | 88-97% | 97-105% | 81-93% |
| EDTA plasma(n=5) | 82-94% | 83-96% | 91-103% | 82-99% |
| Heparin plasma(n=5) | 83-96% | 95-103% | 88-95% | 86-93% |