| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 4000.00 | 2.368 | 2.289 |
| 2000.00 | 1.687 | 1.608 |
| 1000.00 | 1.284 | 1.205 |
| 500.00 | 0.877 | 0.798 |
| 250.00 | 0.543 | 0.464 |
| 125.00 | 0.325 | 0.246 |
| 62.50 | 0.226 | 0.147 |
| 0.00 | 0.079 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 80-93% | 86% |
| EDTA plasma(n=5) | 90-105% | 97% |
| Heparin plasma(n=5) | 80-90% | 85% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 89-103% | 84-94% | 85-103% | 94-105% |
| EDTA plasma(n=5) | 95-104% | 93-102% | 79-94% | 87-98% |
| Heparin plasma(n=5) | 91-103% | 85-97% | 87-96% | 87-90% |