| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 2000.00 | 2.332 | 2.244 |
| 1000.00 | 1.628 | 1.540 |
| 500.00 | 1.243 | 1.155 |
| 250.00 | 0.838 | 0.750 |
| 125.00 | 0.555 | 0.467 |
| 62.50 | 0.354 | 0.266 |
| 31.25 | 0.211 | 0.123 |
| 0.00 | 0.088 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 87-99% | 93% |
| EDTA plasma(n=5) | 87-95% | 91% |
| Heparin plasma(n=5) | 80-95% | 87% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 87-96% | 85-92% | 79-96% | 89-100% |
| EDTA plasma(n=5) | 85-97% | 93-101% | 97-103% | 91-99% |
| Heparin plasma(n=5) | 82-96% | 92-105% | 83-92% | 85-100% |