48 T | $265.00 |
96 T | $376.00 |
Concentration (pg/mL) | OD | Corrected OD |
---|---|---|
1000.00 | 2.078 | 1.984 |
500.00 | 1.489 | 1.395 |
250.00 | 1.267 | 1.173 |
125.00 | 0.872 | 0.778 |
62.50 | 0.521 | 0.427 |
31.25 | 0.294 | 0.200 |
15.63 | 0.211 | 0.117 |
0.00 | 0.094 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 88-105% | 97% |
EDTA plasma(n=5) | 83-100% | 92% |
Heparin plasma(n=5) | 84-114% | 99% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 84-107% | 94-102% | 83-105% | 87-118% |
EDTA plasma(n=5) | 90-97% | 83-101% | 83-112% | 93-101% |
Heparin plasma(n=5) | 92-110% | 90-110% | 86-96% | 81-96% |