| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 500.00 | 2.207 | 2.121 |
| 250.00 | 1.622 | 1.536 |
| 125.00 | 1.277 | 1.191 |
| 62.50 | 0.940 | 0.854 |
| 31.25 | 0.502 | 0.416 |
| 15.63 | 0.347 | 0.261 |
| 7.81 | 0.213 | 0.127 |
| 0.00 | 0.086 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 94-115% | 105% |
| EDTA plasma(n=5) | 92-100% | 96% |
| Heparin plasma(n=5) | 86-116% | 101% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 86-105% | 84-119% | 81-90% | 90-98% |
| EDTA plasma(n=5) | 81-107% | 87-107% | 88-94% | 83-119% |
| Heparin plasma(n=5) | 84-115% | 86-95% | 87-94% | 86-116% |
