| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 2000.00 | 2.379 | 2.293 |
| 1000.00 | 1.593 | 1.507 |
| 500.00 | 1.263 | 1.177 |
| 250.00 | 0.872 | 0.786 |
| 125.00 | 0.543 | 0.457 |
| 62.50 | 0.357 | 0.271 |
| 31.25 | 0.186 | 0.100 |
| 0.00 | 0.086 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 80-95% | 87% |
| EDTA plasma(n=5) | 79-93% | 86% |
| Heparin plasma(n=5) | 82-95% | 88% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 90-101% | 91-99% | 88-97% | 88-104% |
| EDTA plasma(n=5) | 86-97% | 93-102% | 87-98% | 83-97% |
| Heparin plasma(n=5) | 87-96% | 83-101% | 82-95% | 78-102% |