| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 3000.00 | 0.185 | |
| 1500.00 | 0.310 | |
| 750.00 | 0.675 | |
| 375.00 | 0.833 | |
| 187.50 | 1.258 | |
| 93.75 | 1.596 | |
| 46.88 | 2.101 | |
| 0.00 | 2.883 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 85-95% | 90% |
| EDTA plasma(n=5) | 81-103% | 87% |
| Heparin plasma(n=5) | 81-95% | 88% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 82-97% | 85-94% | 90-99% | 82-101% |
| EDTA plasma(n=5) | 88-104% | 87-96% | 85-93% | 82-90% |
| Heparin plasma(n=5) | 91-101% | 87-98% | 89-103% | 82-94% |