| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 2.368 | 2.281 |
| 500.00 | 1.628 | 1.541 |
| 250.00 | 1.226 | 1.139 |
| 125.00 | 0.897 | 0.810 |
| 62.50 | 0.525 | 0.438 |
| 31.25 | 0.256 | 0.169 |
| 15.63 | 0.215 | 0.128 |
| 0.00 | 0.087 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 89-103% | 96% |
| EDTA plasma(n=5) | 82-97% | 89% |
| Heparin plasma(n=5) | 80-95% | 87% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 95-102% | 80-97% | 81-92% | 87-103% |
| EDTA plasma(n=5) | 92-101% | 87-98% | 85-92% | 79-96% |
| Heparin plasma(n=5) | 87-96% | 85-93% | 82-90% | 85-101% |