| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 500.00 | 2.368 | 2.275 |
| 250.00 | 1.645 | 1.552 |
| 125.00 | 1.226 | 1.133 |
| 62.50 | 0.828 | 0.735 |
| 31.25 | 0.492 | 0.399 |
| 15.63 | 0.325 | 0.232 |
| 7.81 | 0.186 | 0.093 |
| 0.00 | 0.093 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 83-97% | 90% |
| EDTA plasma(n=5) | 94-107% | 100% |
| Heparin plasma(n=5) | 90-103% | 96% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 95-102% | 79-97% | 90-99% | 95-103% |
| EDTA plasma(n=5) | 87-96% | 85-97% | 85-94% | 87-96% |
| Heparin plasma(n=5) | 85-96% | 95-104% | 82-90% | 95-104% |