| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 1.990 | 1.908 |
| 500.00 | 1.506 | 1.424 |
| 250.00 | 1.102 | 1.020 |
| 125.00 | 0.837 | 0.755 |
| 62.50 | 0.474 | 0.392 |
| 31.25 | 0.300 | 0.218 |
| 15.63 | 0.170 | 0.088 |
| 0.00 | 0.082 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 94-105% | 98% |
| EDTA plasma(n=5) | 83-97% | 91% |
| Heparin plasma(n=5) | 82-94% | 88% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 86-93% | 93-101% | 86-97% | 90-101% |
| EDTA plasma(n=5) | 91-99% | 83-89% | 88-104% | 97-103% |
| Heparin plasma(n=5) | 85-104% | 83-96% | 95-103% | 88-95% |