| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.368 | 2.282 |
| 5.00 | 2.045 | 1.959 |
| 2.50 | 1.285 | 1.199 |
| 1.25 | 0.812 | 0.726 |
| 0.63 | 0.585 | 0.499 |
| 0.31 | 0.335 | 0.249 |
| 0.16 | 0.207 | 0.121 |
| 0.00 | 0.086 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 81-95% | 88% |
| EDTA plasma(n=5) | 80-95% | 87% |
| Heparin plasma(n=5) | 95-105% | 100% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 88-97% | 88-103% | 86-92% | 91-99% |
| EDTA plasma(n=5) | 89-97% | 85-94% | 79-96% | 85-99% |
| Heparin plasma(n=5) | 79-93% | 94-106% | 83-97% | 85-92% |