| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 2.389 | 2.306 |
| 500.00 | 1.697 | 1.614 |
| 250.00 | 1.243 | 1.160 |
| 125.00 | 0.833 | 0.750 |
| 62.50 | 0.585 | 0.502 |
| 31.25 | 0.305 | 0.222 |
| 15.63 | 0.221 | 0.138 |
| 0.00 | 0.083 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 88-102% | 95% |
| EDTA plasma(n=5) | 89-103% | 96% |
| Heparin plasma(n=5) | 78-92% | 85% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 90-101% | 83-96% | 86-95% | 90-102% |
| EDTA plasma(n=5) | 82-93% | 95-104% | 82-96% | 85-98% |
| Heparin plasma(n=5) | 85-96% | 87-101% | 89-102% | 85-92% |