| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (µg/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.391 | 2.308 |
| 5.00 | 1.694 | 1.611 |
| 2.50 | 1.251 | 1.168 |
| 1.25 | 0.892 | 0.809 |
| 0.63 | 0.556 | 0.473 |
| 0.31 | 0.355 | 0.272 |
| 0.16 | 0.221 | 0.138 |
| 0.00 | 0.083 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 98-104% | 102% |
| EDTA plasma(n=5) | 94-107% | 100% |
| Heparin plasma(n=5) | 80-97% | 88% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 91-99% | 78-96% | 86-92% | 96-105% |
| EDTA plasma(n=5) | 95-103% | 86-92% | 89-97% | 91-103% |
| Heparin plasma(n=5) | 95-102% | 87-103% | 88-95% | 87-95% |