Simian cTnT/TNNT2(Troponin T Type 2, Cardiac) ELISA Kit
One-click to copy product information
ELK11444
Size: Price:
48 T $412.00
96 T $588.00
96 T*5 $2499.00

Overview

Product name: Simian cTnT/TNNT2(Troponin T Type 2, Cardiac) ELISA Kit
Reactivity: Simian
Alternative Names: CMH2; cTnT; TnTc; Cardiomyopathy,Hypertrophic 2; Cardiac muscle troponin T; TNNT2; Troponin T Type 2, Cardiac
Assay Type: Sandwich
Sensitivity: 6.4 pg/mL
Standard: 1000 pg/mL
Detection Range: 15.63-1000 pg/mL
Sample Type: serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids
Assay Length: 3.5h
Research Area: Cardiovascular biology;
Test principle: The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Simian cTnT/TNNT2. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Simian cTnT/TNNT2. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Simian cTnT/TNNT2, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Simian cTnT/TNNT2 in the samples is then determined by comparing the OD of the samples to the standard curve.

Standard curve

Concentration (pg/mL) OD Corrected OD
1000.00 2.332 2.243
500.00 1.579 1.490
250.00 1.096 1.007
125.00 0.682 0.593
62.50 0.593 0.504
31.25 0.336 0.247
15.63 0.225 0.136
0.00 0.089 0.000

Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

Recovery

Matrices listed below were spiked with certain level of recombinant cTnT/TNNT2 and the recovery rates were calculated by comparing the measured value to the expected amount of cTnT/TNNT2 in samples.
Matrix Recovery range Average
serum(n=5) 93-101% 97%
EDTA plasma(n=5) 85-103% 94%
Heparin plasma(n=5) 95-105% 100%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of cTnT/TNNT2 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 95-106% 87-96% 82-90% 92-101%
EDTA plasma(n=5) 85-102% 82-96% 87-98% 78-92%
Heparin plasma(n=5) 98-105% 86-92% 91-103% 89-97%