| 48 T | $265.00 | 
| 96 T | $376.00 | 
| 96 T*5 | $1598.00 | 
| Concentration (ng/mL) | OD | Corrected OD | 
|---|---|---|
| 10.00 | 2.201 | 2.106 | 
| 5.00 | 1.593 | 1.498 | 
| 2.50 | 1.088 | 0.993 | 
| 1.25 | 0.728 | 0.633 | 
| 0.63 | 0.632 | 0.537 | 
| 0.32 | 0.336 | 0.241 | 
| 0.16 | 0.213 | 0.118 | 
| 0.00 | 0.095 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 93-107% | 100% | 
| EDTA plasma(n=5) | 87-99% | 93% | 
| Heparin plasma(n=5) | 87-99% | 93% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 86-98% | 93-106% | 89-97% | 85-94% | 
| EDTA plasma(n=5) | 79-96% | 85-99% | 79-93% | 89-97% | 
| Heparin plasma(n=5) | 85-94% | 87-101% | 89-99% | 85-95% |