| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 4000.00 | 2.363 | 2.282 |
| 2000.00 | 1.597 | 1.516 |
| 1000.00 | 1.029 | 0.948 |
| 500.00 | 0.659 | 0.578 |
| 250.00 | 0.437 | 0.356 |
| 125.00 | 0.256 | 0.175 |
| 62.50 | 0.135 | 0.054 |
| 0.00 | 0.081 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 78-95% | 86% |
| EDTA plasma(n=5) | 95-107% | 101% |
| Heparin plasma(n=5) | 82-95% | 88% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 83-92% | 91-96% | 85-93% | 89-101% |
| EDTA plasma(n=5) | 95-101% | 86-97% | 81-97% | 90-99% |
| Heparin plasma(n=5) | 83-97% | 81-103% | 91-102% | 93-107% |