Products

Mouse 8-OHdG(8-Hydroxydeoxyguanosine) ELISA Kit
One-click to copy product information
ELK11273
Size: Price:
48 T $320.00
96 T $458.00
96 T*5 $1947.00

Overview

Product name: Mouse 8-OHdG(8-Hydroxydeoxyguanosine) ELISA Kit
Reactivity: Mouse
Alternative Names: 8OHdG; 7,8-Dihydro-8-Oxo-2'-Deoxyguanosine; 7,8-Dihydro-8-Oxodeoxyguanosine; 8-Hydroxy-2'-Deoxyguanosine; 8-Oxo-DG; 7,8-dihydrodeoxyguanosine
Assay Type: Competitive Inhibition
Sensitivity: 0.48 ng/mL
Standard: 100 ng/mL
Detection Range: 1.56-100 ng/mL
Sample Type: serum, plasma and other biological fluids
Assay Length: 2h
Research Area: Metabolic pathway;Tumor immunity;Infection immunity;Endocrinology;Hematology;Hepatology;
Test principle: This assay employs the competitive inhibition enzyme immunoassay technique. The microtiter plate provided in this kit has been pre-coated with Mouse 8-OHdG. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Mouse 8-OHdG. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Mouse 8-OHdG in the samples is then determined by comparing the OD of the samples to the standard curve.

Standard curve

Concentration (ng/mL) OD Corrected OD
100.00 0.198
50.00 0.413
25.00 0.597
12.50 0.993
6.25 1.258
3.13 1.522
1.57 1.887
0.00 2.256

Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

Recovery

Matrices listed below were spiked with certain level of recombinant 8-OHdG and the recovery rates were calculated by comparing the measured value to the expected amount of 8-OHdG in samples.
Matrix Recovery range Average
serum(n=5) 87-99% 93%
EDTA plasma(n=5) 80-96% 88%
Heparin plasma(n=5) 92-102% 97%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of 8-OHdG and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 98-105% 82-93% 95-102% 93-101%
EDTA plasma(n=5) 87-96% 82-93% 89-97% 82-90%
Heparin plasma(n=5) 92-101% 95-102% 98-106% 87-98%