| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 200.00 | 1.952 | 1.870 |
| 100.00 | 1.643 | 1.561 |
| 50.00 | 1.067 | 0.985 |
| 25.00 | 0.872 | 0.790 |
| 12.50 | 0.567 | 0.485 |
| 6.25 | 0.324 | 0.242 |
| 3.13 | 0.234 | 0.152 |
| 0.00 | 0.082 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 85-97% | 91% |
| EDTA plasma(n=5) | 87-99% | 93% |
| Heparin plasma(n=5) | 78-93% | 85% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 87-96% | 85-98% | 89-102% | 85-98% |
| EDTA plasma(n=5) | 87-96% | 92-105% | 89-102% | 92-101% |
| Heparin plasma(n=5) | 88-102% | 87-102% | 85-96% | 87-101% |