| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (µg/mL) | OD | Corrected OD |
|---|---|---|
| 200.00 | 2.285 | 2.196 |
| 100.00 | 1.472 | 1.383 |
| 50.00 | 1.135 | 1.046 |
| 25.00 | 0.933 | 0.844 |
| 12.50 | 0.474 | 0.385 |
| 6.25 | 0.361 | 0.272 |
| 3.13 | 0.182 | 0.093 |
| 0.00 | 0.089 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 85-97% | 91% |
| EDTA plasma(n=5) | 79-93% | 86% |
| Heparin plasma(n=5) | 82-95% | 88% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 88-97% | 86-92% | 79-103% | 89-97% |
| EDTA plasma(n=5) | 93-102% | 80-93% | 91-100% | 88-95% |
| Heparin plasma(n=5) | 81-94% | 96-105% | 92-103% | 82-96% |