| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 800.00 | 2.229 | 2.146 |
| 400.00 | 1.607 | 1.524 |
| 200.00 | 0.931 | 0.848 |
| 100.00 | 0.671 | 0.588 |
| 50.00 | 0.476 | 0.393 |
| 25.00 | 0.289 | 0.206 |
| 12.50 | 0.158 | 0.075 |
| 0.00 | 0.083 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 84-113% | 99% |
| EDTA plasma(n=5) | 91-118% | 105% |
| Heparin plasma(n=5) | 84-96% | 90% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 85-116% | 88-115% | 95-107% | 94-119% |
| EDTA plasma(n=5) | 81-118% | 85-111% | 82-99% | 83-109% |
| Heparin plasma(n=5) | 83-103% | 90-100% | 95-120% | 81-95% |