| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (µg/mL) | OD | Corrected OD |
|---|---|---|
| 200.00 | 1.972 | 1.884 |
| 100.00 | 1.593 | 1.505 |
| 50.00 | 1.157 | 1.069 |
| 25.00 | 0.896 | 0.808 |
| 12.50 | 0.543 | 0.455 |
| 6.25 | 0.345 | 0.257 |
| 3.13 | 0.202 | 0.114 |
| 0.00 | 0.088 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 78-90% | 84% |
| EDTA plasma(n=5) | 90-105% | 97% |
| Heparin plasma(n=5) | 95-107% | 101% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 92-105% | 92-101% | 80-94% | 86-101% |
| EDTA plasma(n=5) | 85-94% | 83-96% | 87-98% | 87-100% |
| Heparin plasma(n=5) | 88-99% | 81-93% | 89-97% | 85-94% |