48 T | $412.00 |
96 T | $588.00 |
Concentration (µg/mL) | OD | Corrected OD |
---|---|---|
200.00 | 2.201 | 2.118 |
100.00 | 1.576 | 1.493 |
50.00 | 1.109 | 1.026 |
25.00 | 0.847 | 0.764 |
12.50 | 0.483 | 0.400 |
6.25 | 0.239 | 0.156 |
3.13 | 0.226 | 0.143 |
0.00 | 0.083 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 95-107% | 101% |
EDTA plasma(n=5) | 81-95% | 88% |
Heparin plasma(n=5) | 88-96% | 92% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 85-102% | 83-96% | 95-102% | 87-98% |
EDTA plasma(n=5) | 86-100% | 93-108% | 80-101% | 87-96% |
Heparin plasma(n=5) | 79-96% | 85-99% | 79-93% | 89-97% |