| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 2.045 | 1.972 |
| 500.00 | 1.572 | 1.499 |
| 250.00 | 1.035 | 0.962 |
| 125.00 | 0.728 | 0.655 |
| 62.50 | 0.429 | 0.356 |
| 31.25 | 0.276 | 0.203 |
| 15.63 | 0.148 | 0.075 |
| 0.00 | 0.073 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 92-107% | 99% |
| EDTA plasma(n=5) | 87-99% | 93% |
| Heparin plasma(n=5) | 78-92% | 85% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 88-98% | 87-103% | 89-97% | 85-94% |
| EDTA plasma(n=5) | 95-103% | 79-97% | 93-102% | 87-103% |
| Heparin plasma(n=5) | 93-106% | 87-99% | 85-93% | 79-96% |