| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 2.386 | 2.313 |
| 500.00 | 1.780 | 1.707 |
| 250.00 | 0.929 | 0.856 |
| 125.00 | 0.513 | 0.440 |
| 62.50 | 0.350 | 0.277 |
| 31.25 | 0.240 | 0.167 |
| 15.63 | 0.119 | 0.046 |
| 0.00 | 0.073 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 83-106% | 95% |
| EDTA plasma(n=5) | 82-128% | 105% |
| Heparin plasma(n=5) | 91-117% | 104% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 89-107% | 82-96% | 84-101% | 87-112% |
| EDTA plasma(n=5) | 91-116% | 87-116% | 87-97% | 93-104% |
| Heparin plasma(n=5) | 84-101% | 88-118% | 89-106% | 83-96% |