| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (µg/mL) | OD | Corrected OD |
|---|---|---|
| 200.00 | 2.064 | 1.974 |
| 100.00 | 1.645 | 1.555 |
| 50.00 | 1.188 | 1.098 |
| 25.00 | 0.839 | 0.749 |
| 12.50 | 0.606 | 0.516 |
| 6.25 | 0.349 | 0.259 |
| 3.13 | 0.189 | 0.099 |
| 0.00 | 0.090 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 87-105% | 96% |
| EDTA plasma(n=5) | 85-102% | 94% |
| Heparin plasma(n=5) | 89-97% | 93% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 81-114% | 90-102% | 90-95% | 84-107% |
| EDTA plasma(n=5) | 84-90% | 86-104% | 88-101% | 91-102% |
| Heparin plasma(n=5) | 91-105% | 84-105% | 91-106% | 82-105% |