| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 2.354 | 2.281 |
| 500.00 | 1.790 | 1.717 |
| 250.00 | 0.915 | 0.842 |
| 125.00 | 0.505 | 0.432 |
| 62.50 | 0.340 | 0.267 |
| 31.25 | 0.230 | 0.157 |
| 15.63 | 0.116 | 0.043 |
| 0.00 | 0.073 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 95-107% | 101% |
| EDTA plasma(n=5) | 90-105% | 97% |
| Heparin plasma(n=5) | 78-90% | 84% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 88-97% | 96-103% | 86-92% | 89-97% |
| EDTA plasma(n=5) | 80-101% | 87-96% | 96-105% | 95-103% |
| Heparin plasma(n=5) | 95-103% | 87-98% | 85-92% | 79-96% |