| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 1.946 | 1.857 |
| 500.00 | 1.640 | 1.551 |
| 250.00 | 1.042 | 0.953 |
| 125.00 | 0.686 | 0.597 |
| 62.50 | 0.546 | 0.457 |
| 31.25 | 0.320 | 0.231 |
| 15.63 | 0.198 | 0.109 |
| 0.00 | 0.089 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 95-119% | 107% |
| EDTA plasma(n=5) | 90-112% | 101% |
| Heparin plasma(n=5) | 92-110% | 101% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 83-91% | 83-111% | 94-112% | 94-107% |
| EDTA plasma(n=5) | 91-104% | 94-108% | 82-92% | 86-104% |
| Heparin plasma(n=5) | 85-103% | 87-104% | 91-101% | 86-91% |