| 48 T | $412.00 | 
| 96 T | $588.00 | 
| 96 T*5 | $2499.00 | 
| Concentration (µg/mL) | OD | Corrected OD | 
|---|---|---|
| 20.00 | 2.035 | 1.942 | 
| 10.00 | 1.625 | 1.532 | 
| 5.00 | 1.088 | 0.995 | 
| 2.50 | 0.828 | 0.735 | 
| 1.25 | 0.483 | 0.390 | 
| 0.63 | 0.363 | 0.270 | 
| 0.32 | 0.288 | 0.195 | 
| 0.00 | 0.093 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 82-98% | 90% | 
| EDTA plasma(n=5) | 82-94% | 88% | 
| Heparin plasma(n=5) | 80-97% | 88% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 88-102% | 87-102% | 85-96% | 87-101% | 
| EDTA plasma(n=5) | 88-97% | 86-92% | 95-103% | 98-106% | 
| Heparin plasma(n=5) | 85-94% | 87-96% | 82-95% | 81-93% |