| 48 T | $412.00 | 
| 96 T | $588.00 | 
| 96 T*5 | $2499.00 | 
| Concentration (µg/mL) | OD | Corrected OD | 
|---|---|---|
| 10.00 | 2.132 | 2.046 | 
| 5.00 | 1.515 | 1.429 | 
| 2.50 | 1.142 | 1.056 | 
| 1.25 | 0.864 | 0.778 | 
| 0.63 | 0.563 | 0.477 | 
| 0.32 | 0.308 | 0.222 | 
| 0.16 | 0.256 | 0.170 | 
| 0.00 | 0.086 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 78-96% | 87% | 
| EDTA plasma(n=5) | 86-99% | 92% | 
| Heparin plasma(n=5) | 89-103% | 96% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 88-102% | 86-92% | 81-93% | 89-104% | 
| EDTA plasma(n=5) | 87-96% | 91-101% | 87-98% | 89-103% | 
| Heparin plasma(n=5) | 84-94% | 85-103% | 94-105% | 92-101% |