| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (µg/mL) | OD | Corrected OD |
|---|---|---|
| 100.00 | 2.126 | 2.039 |
| 50.00 | 1.771 | 1.684 |
| 25.00 | 1.098 | 1.011 |
| 12.50 | 0.831 | 0.744 |
| 6.25 | 0.541 | 0.454 |
| 3.13 | 0.352 | 0.265 |
| 1.57 | 0.236 | 0.149 |
| 0.00 | 0.087 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 80-94% | 87% |
| EDTA plasma(n=5) | 91-105% | 98% |
| Heparin plasma(n=5) | 85-97% | 91% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 89-101% | 79-92% | 82-93% | 83-97% |
| EDTA plasma(n=5) | 86-93% | 93-101% | 85-98% | 90-101% |
| Heparin plasma(n=5) | 91-99% | 83-97% | 87-105% | 87-99% |