| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (µg/mL) | OD | Corrected OD |
|---|---|---|
| 200.00 | 2.136 | 2.050 |
| 100.00 | 1.718 | 1.632 |
| 50.00 | 1.168 | 1.082 |
| 25.00 | 0.714 | 0.628 |
| 12.50 | 0.500 | 0.414 |
| 6.25 | 0.272 | 0.186 |
| 3.13 | 0.225 | 0.139 |
| 0.00 | 0.086 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 92-102% | 97% |
| EDTA plasma(n=5) | 93-112% | 103% |
| Heparin plasma(n=5) | 85-92% | 89% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 88-115% | 93-119% | 83-110% | 84-98% |
| EDTA plasma(n=5) | 88-99% | 82-89% | 89-109% | 85-116% |
| Heparin plasma(n=5) | 92-105% | 86-109% | 90-106% | 93-115% |